Performance indicators PRO2/PD03 BR15/BM15&BD30/MC30
Performance front-end pump, built-in front-end pump, external front-end pump, and trolley
Pre pump oil mechanical pump DS-42 2 m3/hr (34l/m) DS-302 14 m3/hr (232l/m)
Vortex dry pump combo-2m:/hr (34l/m) TS-620 30 m/h (500l/m)
The minimum detectable leak rate is 1000ppm 5x10-12atm.cc/sec;
Under helium background, 5x10-12 mbar L/S;5x10-13 Pa m³/mec helium
The maximum pressure resistance of the test port is 13mbar, 10Torr, 1330Pa
The helium pumping speed at the test port is 1.8L/s
Calibration program automatic or manual (built-in or external)
The helium background clearing button resets to zero and has zero point tracking function
High definition color screen and TFT touch screen for operation interface
Language selection: Chinese, English, French, German, Japanese, Korean, Spanish
Loop detection programmable rough sampling time/detection time, set scrap point
Response time<0.5s
Set point standard: 5 set points; Normally open or normally closed: 3 leakage rate points; 1 pressure point; 1 voice point
Serial interface RS-232 relay output interface, 9600 baud (DB-9S)
Relay type I/O output relay output, 5-24V DC output and input (DB-25S)
Wireless remote control (optional) 2.4GHz ISM frequency band, 100 meters (328 feet) range, 4 AA batteries or 5V DC power supply